EĆ16 800 Series Operator's Manual
Level RecoveryExpectedObservedWRSD*nMode
4 Syringe 24 0.97 150.5 149.7 100.5
Capillary 18 0.86 150.2 149.7 100.3
Microcapillary 24 2.00 151.8 149.7 101.4
Microsyringe 24 0.80 150.6 149.7 100.6
Expired Gas 24 0.25 150.4 149.7 100.5
5 Syringe 24 1.20 377.9 377.9 100.0
Capillary 18 2.40 379.0 377.9 100.3
Microcapillary 24 4.08 380.7 377.9 100.7
Microsyringe 24 3.22 378.9 377.9 100.3
Expired Gas 24 0.66 378.7 377.9 100.2
* WRSD = within-run standard deviation