Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
    
NOTE: If your system has a CO-ox module, follow the procedure described for
the appropriate base model. For example, information identified for an 860 also
applies to an 865.
CAUTION: Do not remove or return the sensors to the measurement module
without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.
CAUTION: Do not expose the glucose and lactate biosensors to deproteinizer or
bleach. Replace the biosensors with test/blank sensors (TB4) before cleaning the
sample path. Reinstall the biosensors within 2 hours.
1. Take the appropriate action.
If you have an . . . Then . . .
840 or 850 go to step 2.
860 a. Remove the glucose and lactate biosensors.
b. Install the test/blank glucose and lactate sensors
(TB4).
c. Go to step 2.
CAUTION:ĄĂDo not expose the reference sensor to bleach. Replace the
reference sensor with the test/blank reference sensor (TB5). Do not substitute a
new reference sensor.
2. Replace the reference sensor with a test/blank reference sensor:
a. Remove the reference sensor from the measurement module and set it
aside.
b. Install a test/blank reference sensor (TB5) in the measurement module.
3. Deproteinize the sample path:
a. Prepare the deproteinizer as directed on the package.
b. Select
2 Maintenance and press Enter.
c. Select
1 Deproteinize and press Enter.
d. Invert the deproteinizer vial several times to mix.
e. Insert an aspiration adapter into the sample port and insert the other end
into the deproteinizer.
f. Press
Analyze.
g. Remove the adapter when prompted.
h. Wait 5 minutes for the deproteinizing cycle to finish.
A message box appears prompting you to condition the sensors.
!
!
860
!
2
Menu Code
1