800 Series Operator's Manual
Grasp the tab on the
reference sensor and
pull it up and out of the
measurement module.
Clot-removal
Line
Push the spring-loaded
latch to the right.
d. Connect the measurement module tubing to the latch.
e. Reinstall the reference sensor, ensuring that O-rings are in place.
8. Press the release tab to close the spring-loaded latch.
9. Ensure that the sensors are aligned correctly.
10. Close the measurement module door.
11. Press
Continue.
A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.
12. Allow the system to warm up at least 15 minutes.
13. Press
Yes to perform a two-point calibration.
14. Analyze a minimum of two levels of QC material to verify sensor
performance.