Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
4Ć119
Troubleshooting the System
Problem Probable Cause and Solutions
Meas Module
Temperature Error
The measurement module temperature is outside the
37.0 ± 0.5°C range. The system cannot accept sample analysis
requests.
1. Ensure that the measurement module door is closed.
2. Wait at least 15 minutes for the system to warm up.
3. Perform the Temperature/pAtm Test, as described in
Section 4.
4. If the temperature control is off, press
Reset Control.
5. Verify that the sample ground/temperature sensor is
installed correctly.
6. Verify that the location of the system meets the
specifications for ambient operating temperatures, as
described in Appendix H, Installation.
7. Check the air filter to ensure that the air vent is not
obstructed and that the filter is clean. See Replacing the Air
Filter in Section 3.
8. Verify that the fan is operating.
Meas Module
Temperature Warning
The measurement module temperature is outside the
37.00 ± 0.15°C range. The system can accept sample analysis
requests.
1. Ensure that the measurement module door is closed.
2. Perform the Temperature/pAtm Test, as described in
Section 4, to check the sample temperature.
3. Wait at least 15 minutes for the system to warm up.
4. Verify that the location of the system meets the
specifications for ambient operating temperatures, as
described in Appendix H, Installation.
5. Check the air filter to ensure that the air vent is not
obstructed and that the filter is clean. See Replacing the Air
Filter in Section 3.
6. Verify that the fan is operating.
(Continued)