Use this procedure to control the way QC results are assigned to a QC file.
Table 5-19 describes the options for assigning QC results.
   
 
On The system automatically determines the appropriate QC file
to assign QC sample results, by comparing the QC sample
results for pH and pCO
2
to the ranges entered in setup for
these parameters.
NOTE: If the pH, pCO
2
, or tHb sensor is disabled or not in
calibration, Auto ID does not work.
Off The operator assigns the QC sample results to the appropriate
file. The default value is Off.
1. Access the Global QC Settings screen from the Menu screen:
a. Select 5 Operating Setup and press Enter.
b. Select 1 QC Setup and press Enter.
c. Select
2 Global QC Settings and press Enter.
The Global QC Settings screen appears with the cursor in the Auto ID frame,
as shown in Figure 5-31.
    
    
    
 
2. Select On or Off and press Enter.
3. Press
Done.
4. Define another setup function or press
Exit Menu to return to the Ready
screen.
