Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
1Ć41
Learning About the System
Intended Use of Reagents
The reagents used on the 800 systems have the following intended uses:
Reagent Intended Use
7.3/CO-ox Zero 7.3 reagent provides the calibration point for the one- and
two-point pH and electrolyte calibrations and the zero
calibration for the CO-ox module. 7.3 reagent is buffered to a
pH of 7.382 at 37
°C and is NIST traceable.
6.838 Buffer 6.838 Buffer provides the slope point for the two-point pH and
electrolyte calibrations. 6.838 Buffer is buffered to a pH of
6.838 at 37
°C and is NIST traceable.
Wash G/L Zero Wash G/L Zero cleans the sample path after sample analysis
and during wash sequences. Provides the calibration point for
the two-point metabolite calibration.
Cal G/L Cal G/L provides the calibration point for the one-point
calibration and the slope point two-point metabolite calibration.
Cleaning Solution 1 Cleaning Solution 1, when alternated with Cleaning Solution 2,
provides thorough cleaning of the reagent path.
Cleaning Solution 2 Cleaning Solution 2, when alternated with Cleaning Solution 1,
provides thorough cleaning of the reagent path.
NOTE: The system automatically cleans the reagent path,
using one of these cleaning solutions, every 24 hours at 02:00
(default time). The cleaning cycle is followed by a two-point
calibration.
Conditioner Conditioner cleans and conditions the glass membranes of the
pH and sodium sensors. Conditioner should be used regularly
as part of the preventive maintenance for the system.
CO-oximeter Slope CO-oximeter Slope provides the slope point for the tHb slope
calibration.
Deproteinizer
Deproteinizer removes protein buildup from the sample path.
Deproteinizer should be used regularly as part of the preventive
maintenance for the system.