800 Series Operator's Manual5Ć34
b. Press 2 Maintenance Tasks and press Enter.
The Maintenance Setup screen appears, as shown in Figure 5-19.
   
    
   
 
2. Enter the start date for recording the maintenance tasks.
Tasks become due on midnight of the scheduled date. For example, weekly
tasks are due every 7 days after the start date, monthly tasks are due every
fourth week after the start date.
3. Take the appropriate action.
          
keep the existing
maintenance schedule
go to step 10.
change the frequency of a
maintenance task
go to step 4.
4. Select Edit Frequency.
5. Use the arrow keys to select a task.
6. Select
Select Frequency.
7. Use the arrow keys to select a frequency for the task.
8. Select
OK to change the frequency or select Cancel to return to the
Maintenance Schedule screen.
9. Repeat steps 5 through 8 to change the frequency of another task or go to step
10.
10. Press
Done to return to the Main menu screen.