Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 

   
 
Roll Printer Tests the ability of the roll printer to print all characters.
Bar Code Scanner Tests the ability of the bar code scanner to read a test pattern.
  
Use this procedure to test the ability of the reagent components to deliver reagents
to the fluid detectors in the reagent manifold and the measurement module.
1. Access the Fluidics Functions Test screen from the Menu screen:
a. Select
3 Troubleshooting and press Enter.
b. Select 1 Fluidics System and press Enter.
c. Select 1 Fluidics Functions and press Enter.
The Fluidics Functions Test screen appears as shown in Figure 4-2 for an 860
system.
    
Test one reagent
at a time.
Check here for results.
2. Select the reagent you want to test and press Enter.
3. Press
Start Test.
4. Check the screen for the message, Fluidics functions acceptable, to verify that
the system delivers the reagent to the fluid detectors.
