3Ć22 Rapidlab 800 Operator's Manual
 
Perform the following procedures every 2 months. The bimonthly maintenance
schedule is based on analyzing 30 samples per day, unless otherwise noted. If your
laboratory analyzes more than 30 samples per day, perform bimonthly maintenance
more frequently.
       
Use this procedure to discard and replace the 6.838 Buffer and Wash G/L Zero
reagents if they are opened and installed on the system for 60 days.
Materials required:
S 6.838 Buffer
S Wash G/L Zero
WARNING
Wear safety glasses, gloves, and a laboratory coat when handling the
reagents.
1. At the Ready screen, press
Enter.
The Reagent Levels screen appears.
2. Remove the reagent bottles from the reagent manifold as shown in Figure 3-9
for an 860 system.
        
 
 
3. Write the date installed in the space provided on the new reagent bottles.
NOTE: Do not remove or tighten the cap that contains the reagent septum.
Removing or tightening the cap damages the integrity of the reagent septum.