Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
3Ć100
Rapidlab 800 Operator's Manual
   
Sample Probe
Probe Tab
Slot
Bracket
  
3. Install the new probe:
CAUTION:ĄĂInsert the probe slowly to avoid bending the shaft.
a. Push the sample probe into the opening in the sample port and place the
probe tab in the slot.
b. Connect the sample tubing to the new sample probe.
c. Rotate the sample probe and tubing down toward the system.
d. Place the sample tubing in the bracket on the probe mount.
e. Connect the sample tubing to the sample tubing connector on the
measurement module.
4. Press
Continue.
A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.
5. Press
Yes to perform a two-point calibration.
   
Materials required:
S capillary seal
BIOHAZARD: Refer to Appendix A, Protecting Yourself from Biohazards, for
recommended precautions when working with biohazardous materials.
1. Stop the system from the Menu screen:
!
2
Menu Code
7