3Ć101
Maintaining the System
a. Select 2 Maintenance and press Enter.
b. Select 7 Stop System and press Enter.
2. Ensure that the sample probe is fully retracted.
3. Remove the old capillary seal:
a. Grasp the top of the capillary seal and pull it toward the right and out of the
sample port, as shown in Figure 3-69.
Capillary Seal
Sample Por t
b. Discard the capillary seal according to your laboratory protocol.
4. Install the new capillary seal:
a. Hold the seal at the top, angle it toward the right, and push it into the
opening in the sample port.
b. Insert a capillary tube into the port to ensure that the capillary seal fits
securely in place.
5. Press
Continue.
A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.
6. Press
Yes to perform a two-point calibration.