Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
HĆ12 800 Series Operator's Manual
NOTE:ĄĂIf the temperature is outside of the range, verify that the power has
been on for at least 30 minutes.
40. Access the Barometer
screen from the Menu screen:
a. Select 1 Calibration and press Enter.
b. Select
8 Barometer and press Enter.
The Barometer Calibration screen appears.
41. Compare the displayed atmospheric pressure to the laboratory’s barometer
reading.
If . . . Then . . .
the displayed atmospheric
pressure is correct
press Done. The Ready screen appears.
the displayed atmospheric
pressure is incorrect
a. Type the correct atmospheric pressure, and press
Enter.
b. Press Done to save the new atmospheric pressure
and return to the Ready screen.
c. Perform a gas two-point calibration.
42. Condition the sensors.
Refer to Conditioning the Sensors in Section 3.
CAUTION: Do not remove or return the sensors to the measurement module
without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.
43. Install the glucose and lactate biosensors:
a. Remove
test/blank sensors (TB4).
b. Remove the lactate and glucose biosensors from the foil package.
c. Align the contacts on the biosensors with the contacts in the measurement
module.
d. Snap the body of the biosensors down into place. The contacts must be
flush with the sensors.
e. Press the tab on the spring-loaded latch down to release the latch.
44. Verify that the sensors are installed in the following order:
pO
2
pCO
2
GND pH
Na
+
K
+
Ca
++
Cl
-
RefGlu
Lac
45. Perform three two-point calibrations to start hydrating the glucose and lactate
biosensors as soon as possible.
46. Allow the glucose and lactate biosensors to equilibrate for at least 30 minutes.
47. Verify that the system temperature is within the acceptable range.