3Ć20 Rapidlab 800 Operator's Manual
NOTE: Do not remove or tighten the cap that contains the reagent septum.
Removing or tightening the cap damages the integrity of the reagent septum.
4. Remove the plug from the cap of the new cleaning solution bottle.
5. Insert the cleaning solution bottle into position on the reagent manifold.
6. Push the bottle to ensure that it fits tightly on the reagent fitting.
7. Press
Reset Levels.
The Reset Levels screen appears.
8. Select C1/C2 and press
Done.
      
Use this procedure to discard and replace the 7.3/CO-ox Zero and Cal G/L reagents
if they are opened and installed on the system for 30 days.
Materials required:
S 7.3/CO-ox Zero (7.382 Buffer)
S Cal G/L
WARNING
Wear safety glasses, gloves, and a laboratory coat when handling the
reagents.
1. At the Ready Screen, press
Enter.
The Reagent Levels screen appears.
2. Remove the reagent bottles from the reagent manifold as shown in Figure 3-8
for an 860 system.
       
 
 
