Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
System Administration 5Ć29
Operator IDs can contain 1 to 13 alphanumeric characters and can include spaces.
Operator passwords can contain 1 to 8 alphanumeric characters, can include
spaces, and are case sensitive.
If you enter an operator ID without an operator password, the system does not
accept the entry.
When you have entered 10 operator passwords, press Next Screen to continue
entering passwords. At Screen 50, Next Screen takes you to Screen 1. At Screen 1,
Previous Screen takes you to Screen 50.
      
Use this procedure to define or edit the supervisor password or the password
required to access menu functions.
The supervisor password must be used to access screens in Security Options and
can be used to access sample analysis.
Define a menu options password only if you want to password protect access to
menu functions. Refer to Selecting Password Protection, page 5-26, to select menu
options for password protection.
1. Access the Set Passwords screen from the Menu screen:
a. Select
6 System Setup and press Enter.
b. Select
7 Security Setup and press Enter.
c. Select 1 Set Passwords and press Enter.
NOTE:ĄĂ
You must enter the supervisor password to access the Set Passwords
screen. The default supervisor password is 12345. If you already defined the
supervisor password, you must enter that password.
A message appears prompting you to enter the supervisor password.
2. Type the supervisor password in the Password field and press
OK.
As you type the password, the password does not appear on the screen.
The Set Passwords screen appears, as shown in Figure 5-16.


