Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
DĆ3Connecting to External Devices
f. Press ENTER to confirm your selection.
g. Select the transmission specifications as shown in Table D-1.
  
 
baud rate 9600
parity even
stop bit 1
data bits 8
modem control
no
h. Press ENTER to confirm the selections.
The display shows the port number and the device connected to the port.
i. Press CONTINUE to return to the Select INTERFACES setup option
prompt. The printer prints the port number and the device type connected to
the port.
10. Refer to Defining the Auto Send Option in Section 6 of the
270 CO-oximeter Operator’s Manual to specify whether the 270 automatically
sends results as soon as they are available or whether the operator can select
the results to transmit.
    
Use this procedure to connect an 800 system to the 800 series compatible ticket
printer.
Materials Required:
S ticket printer (part number 477832)
S 800 system to ticket printer interface cable (part number 014116701)
To prevent electrical shock and damage to either system, disconnect
the 800 system and the printer from the AC power source before installing the
cable. Refer to Shutting Down and Restarting the System in Section 5 for the
procedure to disconnect the 800 system.
1. Disconnect the 800 system from the AC power source.
2. Disconnect the printer from the AC power source.
WARNING