Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 

  
1. Analyze the arterial, and the venous or mixed venous samples you want to use
for a-v studies.
2. Select the first patient sample for a-v studies:
a. Select
Menu.
b. Select
4 Data Recall and press Enter.
c. Select 1 Patient Data and press Enter.
The Patient Data Search Criteria screen appears.
3. Enter the patient ID for the sample and press
Done.
The Patient Data Search Log screen appears.
4. Select the patient sample for a-v studies and press
Reporting Options.
The Reporting Options message box appears.
5. Select
Select for aĆv Studies and press OK.
The First a-v Sample Selected message box appears.
6. Press
OK.
If . . . Then . . .
only one sample is found the Patient Data Search Result screen
appears.
Press
Reporting Options. The Reporting
Options message box appears.1
more than one sample is found the Patient Data Search Log screen appears.
a. Select the next sample for a-v studies
and press
Enter.
b. Press
Reporting Options. The
Reporting Options message box appears.
7. Select Select for aĆv Studies and press OK.
The Confirm Samples for a-v Studies message box appears.
NOTE:ĄĂRecord and use the sequence numbers to access the sample results
from the Patient Data Search Log.
8. Enter the Qt value if you want VO
2
and DO
2
results.