3Ć87
Maintaining the System
CAUTION:ĄĂDo not remove or return the sensors to the measurement module
without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.
2. Remove the biosensors as shown in Figure 3-60:
     
         
      
     

a. Push up the latches on the measurement module door and lift the door.
b. Push the spring-loaded latch to the right.
c. Grasp the tab on the biosensor and pull the biosensor up and out of the
measurement module.
d. Discard the biosensor.
3. Remove the new biosensor from the foil package.
4. Ensure that the O-ring is in place on the biosensor.
CAUTION: Do not remove or return the sensors to the measurement module
without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.
NOTE: As you install the biosensors, ensure that the biosensors are in the correct
location. Visually verify that you align the contacts on the biosensors with the
contacts in the measurement module. Slide the remaining sensors to the right to
create more space, if needed.
5. Install the biosensor:
!
!