Operating the System
Message Description
COox Cover Open During
Meas
The cover on the CO-ox module was open while the
system was analyzing the sample.
COox Sample Chamber
Temp Error
The CO-ox module sample chamber temperature is ±
0.35°C of range and cannot accept sample analysis
requests.
COox Sample Temp Out
of Range
The CO-ox module sample chamber temperature is not in
range at the end of a measurement sequence.
Edited Report The Patient Information screen was edited after initial
completion.
Excessive Bubbles in COox
Sample
The CO-ox module detects a non-continuous fluid in the
sample path.
D2 Excessive Drift: __ The sensor drift is beyond predefined limits during a
one-point or a two-point calibration.
Data Entry Incomplete
__ Not Sent
Appears when required data entry fields were not
completed and Auto Send is turned on.
Insufficient COox Sample Appears when there is not enough sample to fill the
CO-ox module sample chamber and the analysis cannot
be performed.
Insufficient Sample Appears when there is not enough sample to fill the
measurement module and you manually positioned the
sample for measurement.
Interfering Substance
Detected
Substances that interfere with analysis are present in the
sample. Only appears when 270 CO-ox is attached.
Interfering Substance: tHb The CO-ox module detects the presence of substances in
the sample that may interfere with the measurement.
Measurement Module
Temp Error
The measurement module temperature is not in range and
the system cannot accept sample analysis requests.
Measurement Module
Temp Warning
The measurement module temperature is ± 0.15°C of
range and can accept sample analysis requests.
__ Not Sent
Appears when you press Do Not Send at the end of
analysis.