Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
  
NOTE: If your system has a CO-ox module, follow the procedure described for
the appropriate base model. For example, information identified for an 860 also
applies to an 865.
1. Access the Condition screen from the Menu screen:
a. Select
2 Maintenance and press Enter.
b. Select
2 Condition and press Enter.
CAUTION: Do not expose the glucose and lactate biosensors to the conditioner.
Replace the biosensors with the test/blank sensors (TB4) before conditioning.
Reinstall the biosensors within 2 hours.
CAUTION: Do not remove or return the sensors to the measurement module
without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.
2. Take the appropriate action.
If you have an . . . Then . . .
840 or 850 go to step 3.
860 a. Remove the glucose and lactate biosensors.
b. Install the test/blank sensors.
c. Go to step 3.
3. Draw the conditioner solution into a syringe.
4. Insert the syringe into the sample port.
5. Press
Analyze.
6. When prompted, remove the syringe.
7. Wait 10 minutes for the conditioning cycle to finish.
A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.
CAUTION: Do not remove or return the sensors to the measurement module
without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.
NOTE: As you reinstall the biosensors, ensure that the biosensors are in the
correct location. Visually verify that you align the contacts on the biosensors with
the contacts in the measurement module. You can slide the remaining sensors to
the right to create more space.
8. Remove the glucose and lactate test/blank sensors and reinstall the glucose and
lactate biosensors as described in Replacing the Glucose and Lactate
Biosensors, page 3-86.