Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
3Ć98
Rapidlab 800 Operator's Manual
   
Materials required:
S sample port
BIOHAZARD: Refer to Appendix A, Protecting Yourself from Biohazards, for
recommended precautions when working with biohazardous materials.
1. Stop the system from the Menu screen:
a. Select
2 Maintenance and press Enter.
b. Select
7 Stop System and press Enter.
2. Pull off the sample door as shown in Figure 3-66.
   
Drip Tray
Sample Door
Retainer Ring
Mount
Sample Port
OĆring
3. Grasp the tab on the retainer ring and firmly pull the tab toward you to rotate
the ring.
4. Grasp the sample port and attached drip tray and pull it off the mount as
shown in Figure 3-66.
NOTE: Ensure that the three O-rings are in place.
5. Install the new sample port on the mount.
6. Push the tab on the retainer ring away from you until it locks into place.
7. Reinstall the sample door, ensuring that it snaps in place.
8. Press
Continue.
A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.
9. Press
Yes to perform a two-point calibration.