Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
3Ć80
Rapidlab 800 Operator's Manual
   


4M KCl Fill Solution
Fill Line
OĆring
OĆring
 
c. Tap the front face of the sensor with your knuckle to release any bubbles.
d. Gradually fill the KCl reservoir with KCl fill solution to the fill line.
CAUTION:ĄĂDo not overtighten the reservoir cap. Overtightening can damage
the gasket and cause leaks.
e. Reinstall the reservoir cap and hand-tighten.
11. Continue with Reinstalling a Sensor, page 3-89.
   
Use this procedure to fill any measurement sensors that do not have sufficient fill
solution. Refer to Table 3-1 to determine the appropriate fill solution to use.
Materials required:
S appropriate fill solution
S lint-free tissue
NOTE: If your system has a CO-ox module, follow the procedure described for
the appropriate base model. For example, information identified for an 860 also
applies to an 865.
NOTE: The pO
2
and pCO
2
sensors do not require fill solution.
NOTE: The glucose and lactate biosensors do not require fill solution.
!
