Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
EĆ48 800 Series Operator's Manual
       
Fetal Hemoglobin at 80% tHb
COHb
MetHb
HHb
–0.9 g/dL
+3.1%
–1.1%
–2.6%
Carboxymethylcellulose tHB
COHb
O
2
Hb
HHb
–2.0 g/dL
+10.0%
–10.0%
+6.5%
* Analytes that are within interference criteria stated in Table EĆ40 are not listed.
  
Quality control materials were analyzed on the 800 series CO-ox module. The
results are presented here.
Precision on aqueous quality control materials was estimated using four 800
systems. One run per instrument was made over a eight day period. Three replicates
of each control level were analyzed in each run.
Table E-43 summarizes the results of the 800 CO-ox module precision on QC
materials.
    
   * 
[
tHb 1 96 9.36 0.05 0.25
2 96 13.98 0.07 0.37
3 96 21.60 0.08 0.54
FO
2
Hb 1 96 5.79 0.11 0.28
2 96 13.56 0.18 0.30
3 96 22.01 0.22 0.30
(Continued)