Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 

   
  
The 800 systems accept QC samples from a syringe or from an aspiration adapter.
When you perform routine QC analysis, all parameters available on your system
are analyzed. You can specify a panel of parameters to analyze for each sample.
BIOHAZARD: Refer to Appendix A, Protecting Yourself from Biohazards, for
recommended precautions when working with biohazardous materials.
1. Press
Change Sample Type.
The Sample Type menu appears as shown in Figure 2-22.
   
   
2. Select Quality Control (QC) and press Enter.
NOTE:ĄĂIf you are using Bayer Diagnostics controls, you can scan the bar code
on the ampule now.
          
analyze all parameters for
this sample
continue with step 3.
specify a panel of
parameters to analyze
a. Select Change Parameters. The Parameter Panel
appears.
b. Select the parameter panel and press
Enter.
c. Continue with step 3.
3. Prepare the sample and insert the device into the sample port. Figure 2-23
shows how to introduce a QC sample using an aspiration adapter.