Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 

  
D60 Port <port> Error:
Problem
The system detects a transmission error in port 1, 2, 3, or the bar code scanner port, which is caused by
a faulty cable connection or by the failure of the communications electronics.
Qualifiers: 1 2 3 4 5
Solutions
Ensure that the cable is firmly connected to the 800 system and to the external
device.
Inspect the cable for wear or crimps.
Perform the procedure to configure the 800 system for the port with the
transmission error. Refer to Configuring for External Devices in Section 5.
Perform the External Loopback Test for the port where the error occurred as
described in External Loopback Test, page 4-75.
If the External Loopback Test fails and the cable is firmly connected to the port,
replace the interface cable.
Perform the External Loopback Test again to ensure that the system sends and
receives the transmissions. If the D code appeared when the system was
transmitting results to the external device connected to the port, you can recall
the results and transmit them again. You can also recall and transmit results
obtained while the port was not operating. Refer to Recalling Patient Sample
Data, Recalling QC Data, or Recalling Calibration Data in Section 2.
If the D code reappears, contact your Service Representative.
D60