Use this procedure to select sample analysis and menu functions for password
protection. When you password protect sample analysis or menu functions,
operators must enter a password to access the protected function.
Select the Sample Analysis option to password protect the system from
unauthorized use. Operators must then enter a valid operator password to analyze
samples and to perform any other system function. You can define passwords for
up to 500 operators. Refer to Defining Operator Passwords, page 5-27, to define
passwords for authorized operators.
Select the menu options you want password protected to prevent unauthorized
changes by operators. Operators must then enter the valid menu options password
to access the protected menus. For example, you can password protect the
Operating Setup menu. Operators must then enter the menu options password to
access the Operating Setup menu. They can still access other menu functions
without a password. Refer to Defining the Menu Options or Supervisor Password,
page 5-29, to define the menu options password.
1. Access the Password Protection screen from the Menu screen:
a. Select
6 System Setup and press Enter.
b. Select 7 Security Setup and press Enter.
c. Select 4 Password Protection and press Enter.
NOTE:ĄĂ
You must enter the supervisor password to access the Password
Protection screen. The default supervisor password is 12345. If you already
defined the supervisor password, you must enter that password.
A message appears prompting you to enter the supervisor password.
2. Type the supervisor password in the Password field and press
OK.
As you type the password, the password does not appear on the screen.
The Password Protection screen appears, as shown in Figure 5-14.
   
Select Sample
Analysis to protect
sample analysis and
system functions.
