800 Series Operator's Manual5Ć56
   
Use this procedure to print a QC setup report for any QC file.
1. Access the QC File screen from the Menu screen:
a. Select
5 Operating Setup and press Enter.
b. Select
1 QC Setup and press Enter.
c. Select 1 QC Files and press Enter.
2. Select the file number that you want to print.
3. Press
Print QC Setup.
The QC Setup report prints. Figure 5-34 shows an example of a QC Setup
Report.
   
.............
.......
....................
.........................
840
850
QC SETUP REPORT APR 12 1994 11:42
SYSTEM 850-1001
GLOBAL QC SETUP
Auto ID is: On
Auto accept QC: Off
QC FILE SETUP
File 1
QC ID: 473842
Level: 2
Lot number: 41561
Expiration date: Dec 30 95
MEAN TARGET RANGE ACTION RANGE
pH 7.419 ( 7.399 - 7.439)( 7.389 - 7.449)
pCO
2
43.4 ( 38.4 - 48.4)( 35.9 - 50.9)
pO
2
103.1 ( 96.1 - 110.1)( 92.6 - 113.6)
Na
+
135.0 ( 130.0 - 140.0)( 127.5 - 142.5)
K
+
4.81 ( 4.31 - 5.31)( 4.06 - 5.56)
Ca
++
1.16 ( 1.06 - 1.26)( 1.01 - 1.31)
Cl
-
102 ( 97 - 107)( 94.5 - 109.5)
Glucose 102.5 ( 97.5 - 107.5)( 87.5 - 117.5)
Lactate 0.92 ( 0.71 - 1.11)( 0.61 - 1.21)
ctHb 14.5 ( 13.5 - 15.5)( 13.0 - 16.0)
FO
2
Hb 1.2 ( 0.7 - 1.7)( 0.4 - 2.0)
FCOHb 95.5 ( 93.5 - 97.5)( 92.5 - 98.5)
FMetHb 0.4 ( 0.2 - 0.6)( 0.1 - 0.7)
FHHb 2.6 ( 2.1 - 3.1)( 1.8 - 3.4)
..............................
....................................................................................................................
.................................
860
These results
appear when an
845, 855, or 865
is interfaced.
..............
..............
....................................................................................
CO-ox
4. Press Menu.
5. You can define another setup function or press
Exit Menu to return to the
Ready screen.
