Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
3Ć83
Maintaining the System
   
Fill Solution
OĆring
Electrode Compartment
7. Insert the internal electrode into the electrode compartment and screw it into
place, ensuring that you do not cross-thread the electrode.
8. Continue with Reinstalling a Sensor, page 3-89.
     
 
If you need to replace the reference sensor, refer to Replacing the Reference
Sensor, page 3-71.
If you need to replace the glucose or lactate biosensors, refer to Replacing the
Glucose and Lactate Biosensors, page 3-86.
Materials required:
S appropriate sensor
BIOHAZARD: Refer to Appendix A, Protecting Yourself from Biohazards, for
recommended precautions when working with biohazardous materials.
NOTE: If your system has a CO-ox module, follow the procedure described for
the appropriate base model. For example, information identified for an 860 also
applies to an 865.
1. Stop the system from the Menu screen:
a. Select
2 Maintenance and press Enter.
b. Select 7 Stop System and press Enter.

2
Menu Code
7