D23 Rgt Manifold: No (Qualifier)
Problem
Fluid detector 3 (FD3), fluid detector 4 (FD4), or both detectors do not detect reagent.
Qualifiers: 7.3 6.8 Wash C1/C2 Cal G/L
Solutions
For the 7.382 Buffer, 6.838 Buffer, Wash G/L Zero and Cal G/L reagents, check
the expiration dates and the levels. Ensure that the reagent bottles are installed
properly and in the correct location. If reagents have expired, are low, or are
incorrectly installed, replace the affected reagents as described in Section 3.
If the D23 . . . Then . . .
disappears perform a two-point calibration.
appears perform the Fluidics Functions Test as described on page 4-56
for the appropriate reagent.
For the C1/C2 reagent, check the expiration date and level. Ensure that the
reagent bottle is installed properly and in the correct location. If the C1/C2 has
expired, is low, or is incorrectly installed, replace it. Prime the system.
Perform the Fluidics Functions Test as described in Fluidics Functions Test,
page 4-56. If the D23 disappears, perform an Auto Clean as described in
Performing the Automatic Clean Sequence in Section 3.
Check the reagent fittings for obstructions and clean the fittings as described in
Cleaning the Reagent Fittings in Section 3.
Perform the Pump Flow Rate Test as described on page 4-57.
Perform the Valves Test as described on page 4-62 to verify that the 6.8, 7.3,
Glu/Lac, Clean, and Vent valves are working.
Perform the Fluid Detector Test as described on page 4-66.
If the D code reappears, contact your Service Representative.
1
Menu Code
1
3
D23