Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
HĆ20 800 Series Operator's Manual
2. Plug the power cord into the AC power receptacle.
At startup, the system proceeds through a series of initial tests and then
displays
Initializing in the banner. Any errors incurred during initialization are
posted to the status log.
3. Access the Service Setup screen from the Menu screen:
a. Select
8 Service Setup and press Enter.
b. Select 4 System ID and press Enter.
4. Type the service password and press
OK.
5. Select the system configuration(s) that you want and press
Enter.
6. Enter the CO-ox module serial number and press
Enter.
7. Press Done.
8. You can define another setup function or press
Exit Menu to return to the
Ready screen.
9. Access the Menu screen and set up your system.
Refer to System Administration in Section 5 for more information about setting
up the system.
10. Initiate a two-point calibration from the Menu screen:
a. Select
1 Calibration and press Enter.
b. Select
2 Two-point and press Enter.
c. Verify acceptable drift.
11. Initiate a tHb slope from the Menu screen:
a. Select
1 Calibration and press Enter.
b. Select
9 tHb Slope and press Enter.
c. Verify acceptable drift.
12. Analyze QC material and verify acceptable results.
Relocating Your System
No special handling or preparation is required to relocate your system to another
area in your laboratory. To store your 800 system for an extended period of time or
to ship it to another location, perform the following procedures, or contact your
Bayer Diagnostics Service Representative for assistance.
Materials required:
S 10% solution of household bleach
S sterile water
S reagent water
8