Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 

   
CAUTION: Do not remove or return the sensors to the measurement module
without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.
2. Remove the sensor as shown in Figure 3-59:
a. Push up the latches on the measurement module door and lift the door.
      
        
      

   
  
b. Push the spring-loaded latch to the right.
c. Grasp the tab on the sensor and pull the sensor up and out of the
measurement module.
d. Discard the sensor.
3. Perform the appropriate action.
          
the chloride sensor go to step 4.
other measurement sensors or the
sample ground/temperature sensor
go to step 5.
4. Fill the chloride sensor with the Na
+
, K
+
, Cl
, Ca
++
fill solution:
a. Remove the new sensor from the box.
CAUTION:ĄĂDo not touch the internal electrode wire. The wire is fragile and is
easily damaged.
b. Unscrew the internal electrode and set it aside on a lint-free tissue.
c. Rinse the electrode compartment with 3 drops of fill solution and then empty
it.
d. Fill the sensor almost full, leaving a bubble at the top.
!
860850
!