Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
800 Series Operator's Manual5Ć32
2. Change the date and time.
   
        
Date a. Move the cursor to the Date field.
b. Type the month, date, and year in the selected format. Press
Enter after you complete the field. The cursor moves to the
Time field.
Time a. Move the cursor to the Time field.
b. Type the hour and minutes in the format, HH:MM. You can
enter 0 – 23 hours and 0 – 59 minutes.
c. Press
Enter after you complete the field.
Date Format
a. Move the cursor to the required format.
b. Press
Enter. The date is reformatted, if necessary.
3. Press Done when you finish.
If you changed the date or format, the Updating Database message box
appears.
4. Define another setup function or press
Exit Menu to return to the Ready
screen.
If you press Done before completing all required fields, the Incomplete Data Entry
message box appears. Press
OK and complete the required fields.
If you type an invalid value in a field and press Done, the Invalid Entry message
box appears. Press
OK and type a valid entry in the field.
To type a single number in a date or time field, you must precede the number with
a zero. For example, to enter June as the date, you must enter 06 in the month field.
 
Use this procedure to select the language that the system uses to present
information on the screen and in printed reports.
1. Access the Service Setup screen from the Menu screen:
a. Select
8 Service Setup and press Enter.
b. Select 3 Language Selection and press Enter.
2. Select the appropriate language and press
Enter.
3. Press
Done.
4. You can define another setup function or press
Exit Menu to return to the
Ready screen.


