Rapidlab 800 Operator's Manual
The following table describes the messages that may appear in patient sample
reports on the roll printer:
Message Description
****** = Not in
Calibration
The sensor is out of calibration.
------ or ------ = Out of
Range: ____
The result is above or below the measurement range.
* = D5 No Endpoint: __ The sensor did not reach a stable reading within the
predefined time limit.
* = exception or D–code
noted (in Patient Data Log)
The sample has an exception or D code. Appears in
Patient Data Log.
or = exceeds reference
range
The result is above or below the reference range.
↑↑
or ↓↓ = exceeds action
range
The result is above or below the action range.
? = If blood, question data Optical measurements indicate that the CO-oximeter
results should be reviewed. See Troubleshooting Patient
Results in Section 4.
SulfHb > 1.5% The system detects a sulfhemoglobin concentration
greater than 1.5% in a CO-ox sample.
# = Interfering Substance:
Glu
The system detects the presence of substances in the
sample that may interfere with glucose measurement.
NOTE: Repeated, unexpected occurrence of this
message may indicate sensor failure. See Troubleshooting
Patient Results in Section 4.
# = Interfering Substance:
Lac
The system detects the presence of substances in the
sample that may interfere with lactate measurement.
NOTE: Repeated, unexpected occurrence of this
message may indicate sensor failure. See Troubleshooting
Patient Results in Section 4.
D38 Temperature Error The system detects an error in the temperature control
system of the base model.
Bubbles Detected In
Sample
The system detects a non-continuous fluid in the sample
path of the base model.
(Continued)