Troubleshooting the System
BIOHAZARD: Refer to Appendix A, Protecting Yourself from Biohazards, for
recommended precautions when working with biohazardous materials.
NOTE: Do not use a needle on the syringe that you use to introduce reagent water
or air into the CO-ox sample tubing. Using a needle may damage the CO-ox
sample tubing.
1. Stop the system from the Menu screen:
a. Select
2 Maintenance and press Enter.
b. Select 7 Stop System and press Enter.
2. Inspect the CO-ox sample tubing and waste tubing for obstructions, as shown
in Figure 4-20.
Sample Chamber
CO-ox Pump
Waste Tubing
Reagent Pump
Hemolyzer
FD5
Sample
Connector
Inlet
Outlet
Bubble Trap
2
Menu Code
7