Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
DĆ6 800 Series Operator's Manual
Connecting to a Laboratory or Hospital Information
System
Use this procedure to connect an 800 system to a laboratory information system
(LIS) or hospital information system (HIS).
Materials Required:
S LIS or HIS to 800 system interface cable and 9-pin, cable-matching connector
(supplied with the LIS or HIS)
S if you are converting from a 200 series system, 9-pin to 25-pin cable adapter
(part number 014118001)
To prevent electrical shock and damage to either system, disconnect
the 800 system and Data Management system from the AC power source before
installing the cable. Refer to Shutting Down and Restarting the System in Section 5
for the procedure to disconnect the 800 system.
1. Contact your LIS or HIS system manager to determine the appropriate
interface cable.
2. Disconnect the 800 system from the AC power source.
3. Connect the interface cable to serial port 2 or 3 on the 800 system and to the
LIS or HIS. Refer to Figure D-1.
4. Tighten the hold down screws on the connectors.
5. Restore power to the 800 system.
6. Refer to Configuring for External Devices in Section 5 to configure the 800
system for the LIS or HIS.
7. Refer to the 800 Series Interface Specification Manual to configure the LIS or
HIS for the 800 system.
WARNING