5Ć6 800 Series Operator's Manual
1. Access the System Information screen from the Menu screen:
a. Press
8 Service Setup and press Enter.
b. Press
1 System Information and press Enter.
The System Information screen appears, as shown in Figure 5-1.
      
    
  
  
 


2. Press Menu when you finish viewing the screen.
3. Define another setup function or press
Exit Menu to return to the Ready screen.
    
Use this procedure to select the fields you want to use in the Patient Information
screen. The Patient Information screen can contain 16 fields. The default for each
field is On, which means that all fields appear on the Patient Information screen.
You can choose to have a field not appear on the form by turning the field off.
Certain fields can be made required fields. Required fields are fields that must be
completed to continue with analysis. You can make a field required by selecting the
check box next to the field. The default for required fields is Off, which means that
none of the fields are required. Table 5-2 describes the Patient Information fields.
  
 
Patient ID number used to identify the patient
Accession # number used to identify the sample
Location
location of the patient