800 Series Operator's Manual5Ć40
      
   
  
   
2. Select High or Low and press Enter.
3. Press
Done.
4. Define another setup function or press
Exit Menu to return to the Ready
screen.
The option that you select for the Reporting Resolution is saved with each patient
and QC sample. If you recall the sample data later and have changed the
resolution, the system prints the report using the resolution selected when the
sample was saved.
    
Use this procedure to define the way the system processes patient sample results
when a sensor is out of calibration. When the system detects a sensor that has
excessive calibration drift, it cannot analyze samples until the drift is corrected. By
selecting Yes for Report Results with Cal Drift, you have the option to let the
system analyze a sample when a sensor is out of calibration.
     
 
Yes The system analyzes the sample and logs the event in the status log. The
printed report contains the message, not in calibration.
No A D2 code appears and the affected sensor is turned off. Analysis continues
but no results are generated for the affected sensor. The default value is No.

