Rapidlab 800 Operator's Manual
9. Perform a two-point calibration.
a. Press Calibrate.
b. Select Two-point and press Enter.
c. Press
Start Calibration.
Perform two-point calibrations after changing the calibration reagents to ensure
that the reagents are acceptable and the system is functioning properly. You can
analyze quality control materials with the new reagents and compare the results
with the previous QC results after the system is recalibrated.
Replacing the Printer Paper
CAUTION: Do not attempt to print without paper installed or with a double
thickness of paper in the printer. Use the paper advance key to feed the paper
through the printer.
1. Stop the system from the Menu screen:
a. Select
2 Maintenance and press Enter.
b. Select 7 Stop System and press Enter.
2. Pull up the paper spool.
3. Firmly grasp the old roll of paper and pull it off the spool.
NOTE: Grasp the roll of paper tightly to slide the paper completely off the spool.
4. Reinstall the paper spool.
5. Lift the printer cover.
6. Push up the printer lever.
7. Install a roll of printer paper:
a. Unroll a small amount of paper from the new roll and cut or fold a clean
straight edge, if necessary.
b. Place the paper roll in the cavity with the paper unrolling from the bottom
as shown in Figure 3-38.