System Administration 5Ć61
c. Select 2 Cal Intervals and press Enter.
The Cal Intervals screen appears with the cursor in the Calibration Frequency
frame, as shown in Figure 5-37.
      
  
   
   
  
  
  
    
  

2. Select the options you want.
           
Calibration Frequency Select Fixed Time or Flexible Time and press Enter.
NOTE: If you select Flexible Time the interval and
Auto Repeat options are not available.
One-point, Two-point, or
tHb Slope Interval
a. Move to the appropriate field.
b. Type the amount of time you want between
calibrations and press
Enter.
Auto Repeat a. Move to the Auto Repeat frame.
b. Select
On and press Enter to automatically repeat
calibrations.
Select
Off and press Enter to stop repeat automatic
calibrations.
Metabolite Recal
a. Move to the Metabolite Recal frame.
b. Select
On and press Enter to perform one-point
metabolite calibrations automatically.
Select
Off and press Enter to stop automatic
one-point metabolite calibrations.
3. Press Done when you finish.
4. You can define another setup function or press
Exit Menu to return to the
Ready screen.