Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
Installation HĆ3
 
 
 
 

 

 
  
 
  

  
  
  
13. Rotate the system so that the right side panel faces you.
14. Install the fuses as required for your system
       
your system has no voltage bobbin,
it uses a universal power supply
suitable for 100V to 240V
gently pry open the fuse compartment door
using a small, flat–blade screwdriver and
continue with step e.
your system has a voltage bobbin
continue with step a.
a. Take the voltage bobbin from the customizing box and the fuse kit from the
installation kit.
b. Gently pry the fuse compartment door open using a small, flat-blade
screwdriver.
c. Select the operating voltage.
Rotate the voltage bobbin so that the correct voltage faces you.
d. Install the voltage bobbin as shown in Figure H-2.