Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 

   
This flag . . . Indicates . . .
the result is below the lower limit of the target range.
↑↑ the result is above the upper limit of the action range.
↓↓ the result is below the lower limit of the action range.
- - - the result is above the upper limit of the measurement range.
- - -
the result is below the lower limit of the measurement range.
The following table describes the messages that may appear in QC sample reports
on the roll printer:
Message Description
****** = Not in Calibration Sensor is out of calibration.
* = D5 No Endpoint: __ QC sample analysis did not reach endpoint.
or = exceeds target range QC sample result is above or below the target range.
↑↑
or ↓↓ = exceeds action
range
QC sample result is above or below the action range.
ACCEPTED INTO QC FILE
__
Indicates the QC result status (accepted) and the file
number in which the results are stored.
Bubbles Detected In Sample System detects a non-continuous fluid in the
measurement module sample path.
D2 Excessive Drift: __ Sensor drift is beyond predefined limits during a
one-point or a two-point calibration.
D38 Temperature Error System detects an error in the temperature control
system.
DISCARDED INTO QC
FILE __
Indicates the QC result status (discarded) and the file
number in which the results are stored.
REJECTED INTO QC
FILE __
Indicates the QC result status (rejected) and the file
number in which the results are stored.