System Administration 5Ć15
If you change parameter units for pH after the system has started to collect data,
QC statistics are still computed in the current units.
  
Use this procedure to define the values for ctHb and p50 that the system uses when
no value is entered or measured, and to define the value for the O
2
binding factor
and ctO
2
(a-v), which are constants supplied by the system. The system uses these
values for certain calculations. Table 5-5 lists the default values and valid ranges
for ctHb, O
2
binding factor, p50, and ctO
2
(a-v).
  
     
ctHb 2.0 – 25.0
20 – 250
1.2 – 15.5
15.0 g/dL
g/L
mmol/L
p50 0.0 – 100.0
0.00 – 13.32
26.5 mmHg
kPa
O
2
Binding 1.35 – 1.39 1.39
ctO
2
(a-v) 0.0 – 20.0
0 – 200
0.0 – 9.0
3.5 mL/dL
mL/L
mmol/L
1. Access the Units and Values screen from the Menu screen:
a. Select
5 Operating Setup and press Enter.
b. Select
4 Units/Values and press Enter.
The first Units and Values screen appears.
2. Press
Next Screen.
The second Units and Values screen appears as shown in Figure 5-7.


5
Menu Code
4