Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
800 Series Operator's Manual5Ć70
   
Use this procedure to view QC data from archive diskettes. You can also print the
following QC reports from archive diskettes:
S Levey-Jennings Chart
S QC Sample Report
S QC Statistical Summary
NOTE: You cannot edit or restore archived QC data.
1. Access the View Archived QC Data screen from the Menu screen:
a. Select
7 System Utilities and press Enter.
b. Select 2 Disk Utilities and press Enter.
c. Select 3 View Archive and press Enter.
The View Archived QC Data screen appears, with the Begin Viewing Archive
message box.
2. Insert the archived diskette in the diskette drive and press
Continue.
The system reads the diskette. When the system is ready, a message appears
containing archive diskette information.
3. Press
Continue.
4. Remove the diskette when the system prompts you. The Archived QC Search
Criteria screen appears.
5. Type the search criteria and press
Enter after you complete each field. Refer to
Recalling QC Data in Section 2, for information about entering the search
criteria.
6. Press
Done when you finish entering your search parameters.
       
more than one QC
sample is found
the Archived QC Search Log appears. The log contains the
reports that the system located for the search criteria you
entered. Continue with step 7.
one QC sample is
found
the Archived QC Search Result screen appears. Continue
with step 8.
