Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
HĆ1
  
Your 800 system should be installed by an authorized Bayer Diagnostics
representative. Use the following procedure to install your 800 system yourself
only if you are located in a region where Bayer Diagnostics Field Service
Representatives do not perform installation. For detailed information about the
software, refer to Learning About the System in Section 1 and Operating the
System in Section 2.
Select a location for your system. Place the 800 system in a location that is not
exposed to direct sunlight. Refer to Table H-1 for the system specifications.
 
 
ambient operating
temperature
15 – 32°C
ambient operating
relative humidity
5 – 85%, non-condensing
power rating 400VA (maximum)
voltage requirements 100V/120V (85V to 132V)
50/60Hz
220V/240V (170V to 264V)
50/60Hz
ambient operating
barometric pressure
400 – 825 mmHg (53.0 – 110.0 kPa)
system dimensions
height 50.8 cm (20 inches)
width 55.9 cm (22 inches)
depth 50.8 cm (20 inches)
weight 29.5 kg (65 lbs)
Materials required:
S Phillips screwdriver
CAUTION: Do not remove the glucose or lactate biosensors from their packages
during room temperature equilibration.
1. Remove the glucose and lactate biosensors from the refrigerator and allow
them to equilibrate at room temperature (18 – 25°C) for at least 1 hour
before use.
!
