Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 

   
Storing Reagents
This section describes the storage instructions for the reagents used on the 800
systems.
CAUTION: Discard any reagent that is frozen. Do not thaw and use. The reagent
composition is irreversibly altered when the reagent is frozen.
NOTE: Do not use any reagents beyond the expiration date. The expiration date
for Bayer Diagnostics reagents indicates the last month of valid use for unopened
reagents.
Reagent Storage Instructions
7.3/CO-ox Zero Store 7.3/CO-ox Zero upright at 4 to 25°C.
Discard 7.3/CO-ox Zero 30 days after opening.
6.838 Buffer Store 6.838 Buffer upright at 4 to 25°C.
Discard 6.838 Buffer 60 days after opening.
Wash G/L Zero Store Wash G/L Zero upright at 4 to 25
°C.
Discard Wash G/L Zero 60 days after opening.
Cal G/L Store Cal G/L upright at 4 to 25
°C.
Discard Cal G/L 30 days after opening.
Cleaning Solution 1 Store Cleaning Solution 1 in the dark. The solution is light
sensitive.
Store Cleaning Solution 1 upright at 4 to 25
°C.
Discard Cleaning Solution 1 four weeks after opening.
Cleaning Solution 2 Store Cleaning Solution 2 in the dark. The solution is light
sensitive.
Store Cleaning Solution 2 upright at 4 to 25
°C.
Discard Cleaning Solution 2 four weeks after opening.
Conditioner Store Conditioner at 4 to 25
°C.
Discard Conditioner 24 hours after opening.
Deproteinizer
Store Deproteinizer at 4 to 25°C. Activated material is stable for
24 hours when stored at 4 to 8
°C.
Discard Deproteinizer 24 hours after preparing.
!