Measurement Performance
There are two versions of the meter verification application: the original version and Micro Motion
Smart Meter Verification. Table 10-1 lists requirements for each version. Table 10-2 provides a
comparison of the two versions.
Note: If you are running an older version of ProLink II or the EDD, you will not be able to access the
additional features in Smart Meter Verification. If you are running an updated version of ProLink II
or the EDD with the original version of meter verification, the meter verification procedures will be
slightly different from the procedures shown here.
Table 10-1 Version requirements for meter verification application
Requirement type
Meter verification application
Original version Smart Meter Verification
Transmitter v1.0 v1.4
ProLink II requirements v2.5 v2.9
EDD requirements 2400SDP_pdmrev1_00 folder 2400SDP_pdmrev1_40 folder
Table 10-2 Comparison of meter verification features and functions: original version vs. Smart Meter
Verification
Feature or function
Meter verification application
Original version Smart Meter Verification
Process interruption No need to halt flow No need to halt flow
Measurement interruption Three minutes. Outputs go to:
• Last Measured Value
• Configured Fault Value
User option:
• Continue Measurement. Measurement is
not interrupted. Test requires
approximately 90 seconds.
• Last Measured Value. Outputs fixed and
measurement interrupted for
approximately 140 seconds.
• Configured Fault Value Outputs fixed and
measurement interrupted for
approximately 140 seconds.
Result storage Test results stored only for tests run with
ProLink II, and stored on the PC
Twenty most recent results stored on the
transmitter, independent of tool used to
perform the procedure. For tests run with
ProLink II, additional result data stored on
PC.
Result data on display Pass/Fail/Abort for current test For all results stored on transmitter:
• Pass/Fail/Abort
• Abort code (if relevant)
• Stiffness of the right and left pickoffs