Model 2400S Transmitters for PROFIBUS-DP
PROFIBUS Bus Parameters
14 SNS_StatusWords5 B_ENUM D (20 Hz) R • 0x0001 = Boot sector (CP)
• 0x0002 = Not Used
• 0x0004 = Not Used
• 0x0008 = Not Used
• 0x0010 = Not Used
• 0x0020 = Not Used
• 0x0040 = D3 calibration in progress
• 0x0080 = D4 calibration in progress
• 0x0100 = Not used
• 0x0200 = Not used
• 0x0400 = Temperature slope calibration in
progress
• 0x0800 = Temperature offset calibration in
progress
• 0x1000 = FD calibration in progress
• 0x2000 = D2 calibration in progress
• 0x4000 = D1 calibration in progress
• 0x8000 = Zero calibration in progress
15 SNS_StatusWords6 B_ENUM D (20 Hz) R • 0x0001 = Not Used
• 0x0002 = Not Used
• 0x0004 = Not Used
• 0x0008 = Not Used
• 0x0010 = Not Used
• 0x0020 = Not Used
• 0x0040 = Not Used
• 0x0080 = Not Used
• 0x0100 = DE
0
active
• 0x0200 = DE
1
active
• 0x0400 = DE
2
active
• 0x0800 = DE
3
active
• 0x1000 = DE
4
active
• 0x2000 = Not Used
• 0x4000 = Not Used
• 0x8000 = Incorrect board type (A030)
16 SNS_StatusWords7 B_ENUM D (20 Hz) R 0x0001 = K1/FCF combination unrecognized
• 0x0002 = Warming up
• 0x0004 = Low power (A031)
• 0x0008 = Tube not full (A033)
• 0x0010 = Meter verification / Outputs in fault
(A032)
(1)
• 0x0020 = Meter verification / Outputs at last
value (A131)
(1)
• 0x0040 = PIC UI EEPROM Error (A133)
• 0x0080 = Not Used
• 0x0100 = Not Used
• 0x0200 = Not Used
• 0x0400 = Not Used
• 0x0800 = Not Used
• 0x1000 = Not Used
• 0x2000 = Not Used
• 0x4000 = Meter verification failed (A034)
• 0x8000 = Meter verification aborted (A035)
Table D-4 Diagnostic block (Slot 3) continued
Index Name
Data
type
Memory
class Access Definition/Code/Comments