21 SNS_D1 FLOAT S R/W Density of D1 calibration fluid
22 SNS_D2 FLOAT S R/W Density of D2 calibration fluid
23 SNS_CalValForFD FLOAT S R/W Density of flowing density calibration fluid
24 SNS_TseriesD3 FLOAT S R/W Density of D3 calibration fluid
25 SNS_TseriesD4 FLOAT S R/W Density of D4 calibration fluid
26 SNS_DensityTempCoeff FLOAT S R/W Density temperature coefficient
27 SNS_TSeriesFlow
TGCO
FLOAT S R/W T-Series FTG value
28 SNS_TSeriesFlow
FQCO
FLOAT S R/W T-Series FFQ value
29 SNS_TSeriesDens
TGCO
FLOAT S R/W T-Series DTG value
30 SNS_TSeriesDens
FQCO1
FLOAT S R/W T-Series DFQ1 value
31 SNS_TSeriesDens
FQCO2
FLOAT S R/W T-Series DFQ2 value
32 SNS_TempCalOffset FLOAT S R/W Temperature calibration offset
33 SNS_TempCalSlope FLOAT S R/W Temperature calibration slope
34 SNS_EnableExtTemp USINT16 S R/W Use external temperature for API and ED:
• 0x0000 = Disabled
• 0x0001 = Enabled
35 SNS_ExternalTempInput FLOAT S R/W External temperature value
36 SNS_EnablePresComp Method S R/W Pressure compensation:
• 0x0000 = Disabled
• 0x0001 = Enabled
37 SNS_ExternalPresInput FLOAT D (20) R/W Exernal pressure value
38 SNS_PressureUnits USINT16 S R/W Pressure measurement unit
See Table 6-7 for codes
39 SNS_FlowPresComp FLOAT S R/W Pressure correction factor for flow
40 SNS_DensPresComp FLOAT S R/W Pressure correction factor for density
41 SNS_FlowCalPres FLOAT S R/W Flow calibration pressure
42 SNS_FlowZeroRestore S R/W Restore factory zero:
• 0x0000 = No action
• 0x0001 = Restore
43 DB_SNS_AutoZero
Factory
S R Factory value for flow signal offset at zero flow,
in µsec
Table D-3 Calibration block (Slot 2) continued