108 Micro Motion
®
Model 2400S Transmitters for PROFIBUS-DP
Measurement Performance
Detailed test data with PROFIBUS bus parameters
Note: Requires Smart Meter Verification. No detailed test data is available with the original version
of the meter verification application.
For each Smart Meter Verification test, the following data is stored on the transmitter:
Powered-on hours at the time of the test
Test result
Stiffness of the left and right pickoffs, shown as percentage variation from the factory value. If
the test aborted, 0 is stored for these values.
Abort code, if applicable
To view this data, see Figure 10-13.
Figure 10-13 Meter verification test data – PROFIBUS bus parameters
Table 10-6 PROFIBUS bus parameters test data interface for Smart Meter Verification
Step number Step description Interface
(1)
(1) For detailed information, see Table D-4.
1 Set index Diagnostic block (Slot 3)
Index 87
2 Read test counter Diagnostic block (Slot 3)
Index 88
3 Read test start time Diagnostic block (Slot 3)
Index 89
4 Read test result Diagnostic block (Slot 3)
Index 90
5 Read LPO stiffness Diagnostic block (Slot 3)
Index 91
6 Read RPO stiffness Diagnostic block (Slot 3)
Index 92
Step 3
Read test start time
Step 2
Read test counter
Step 4
Read test result
Step 5
Read LPO stiffness
Step 6
Read LPO stiffness
Step 1
Set index
See Table 10-6.