Emerson Process Management 2400S Pacemaker User Manual


 
Configuration and Use Manual 103
Measurement Performance
Measurement Performance DefaultsTroubleshootingCompensation
Detailed test data with ProLink II
For each test, the following data is stored on the transmitter:
Powered-on hours at the time of the test (Smart Meter Verification)
Test result
Stiffness of the left and right pickoffs, shown as percentage variation from the factory value. If
the test aborted, 0 is stored for these values.
Abort code, if applicable
ProLink II stores additional descriptive information for each test in a database on the local PC,
including:
Timestamp from the PC clock
Current flowmeter identification data
Current flow and density configuration parameters
Current zero values
Current process values for mass flow rate, volume flow rate, density, temperature, and external
pressure
(Optional) User-entered customer and test descriptions
If you are using Smart Meter Verification and you run a meter verification test from ProLink II,
ProLink II first checks for new test results on the transmitter and synchronizes the local database if
required. During this step, ProLink II displays the following message:
Synchronizing x out of y
Please wait
Note: If you request an action while synchronization is in process, ProLink II displays a message
asking whether or not you want to complete synchronization. If you choose No, the ProLink II
database may not include the latest test results from the transmitter.
Table 10-5 Meter verification abort codes
Abort code Description Suggested action
1 User-initiated abort None required. Wait for 15 seconds before
starting another test.
3 Frequency drift Ensure that temperature, flow, and density are
stable, and rerun the test.
5 High drive gain Ensure that flow is stable, minimize entrained
gas, and rerun the test.
8 Unstable flow Review the suggestions for stable flow in
Section 10.3.1 and rerun the test.
13 No factory reference data for meter verification
test performed on air
Contact Micro Motion customer service and
provide the abort code.
14 No factory reference data for meter verification
test performed on water
Contact Micro Motion customer service and
provide the abort code.
15 No configuration data for meter verification Contact Micro Motion customer service and
provide the abort code.
Other General abort. Repeat the test. If the test aborts again, contact
Micro Motion customer service and provide the
abort code.