Emerson Process Management 2400S Pacemaker User Manual


 
114 Micro Motion
®
Model 2400S Transmitters for PROFIBUS-DP
Measurement Performance
Note the following:
If the transmitter was ordered with a display:
- The zero button is not available.
- If the off-line menu has been disabled, you will not be able to zero the transmitter with the
display. For information about enabling and disabling the off-line menu, see Section 8.9.5.
- You cannot change the zero time with the display. If you need to change the zero time, you
must use ProLink II or PROFIBUS protocol.
If the transmitter was ordered without a display, the zero button is available.
- You cannot change the zero time with the zero button. If you need to change the zero time,
you must use ProLink II or PROFIBUS protocol.
- The zero button is located on the user interface board, beneath the transmitter housing
cover (see Figure 3-1). For instructions on removing the transmitter housing cover, see
Section 3.3.
- To press the zero button, use a fine-pointed object that will fit into the opening (0.14 in or
3.5 mm). Hold the button down until the status LED on the user interface module begins to
flash yellow.
During the zero procedure, the status LED on the user interface module flashes yellow.
Figure 10-17 Zero button – Flowmeter zero procedure
Status LED
Status LED flashes
yellow
Done
Solid Green or
Solid Yellow
Troubleshoot
Solid
Red
Press ZERO button